The CLIA compliance calendar
Daily: QC for each test system on each day of patient testing, plus temperature and maintenance logs. Every six months: calibration verification and method comparisons. Twice a year: accuracy checks for tests without regulated proficiency testing. Yearly: competency after an employee's first year. Every two years: CLIA fees and survey.
Calendar
| Frequency | Task | Guide |
|---|---|---|
| Each day of testing | QC (default: two levels quantitative, positive and negative qualitative), unless an IQCP applies | QC |
| Daily or per manufacturer | Maintenance, function checks, temperatures | Maintenance |
| Each PT event | Test, attest, review and investigate PT | PT |
| Every 6 months | Calibration verification | Calibration |
| Twice a year | Compare methods or sites; verify accuracy of non-PT tests | Comparisons |
| Twice in year one, then yearly | Competency assessment | Competency |
| Ongoing, reviewed on a schedule | Quality assessment | QA |
| Every two years | CLIA fees; state or accreditor survey | Fees |
Regulations behind this
| Regulation | What it covers |
|---|---|
| 42 CFR § 493.1256Standard: Control procedures | Control procedures |
| 42 CFR § 493.1255Standard: Calibration and calibration verification procedures | Calibration and calibration verification |
| 42 CFR § 493.1281Standard: Comparison of test results | Comparison of test results |
| 42 CFR § 493.1236Standard: Evaluation of proficiency testing performance | Evaluation of proficiency testing performance |
| 42 CFR § 493.1235Standard: Personnel competency assessment policies | Personnel competency assessment policies |
Full text of the regulations
Tap a regulation to read the complete official text.
§ 493.1256 Standard: Control procedures.
(a) For each test system, the laboratory is responsible for having control procedures that monitor the accuracy and precision of the complete analytic process.
(b) The laboratory must establish the number, type, and frequency of testing control materials using, if applicable, the performance specifications verified or established by the laboratory as specified in § 493.1253(b)(3).
(c) The control procedures must—
(1) Detect immediate errors that occur due to test system failure, adverse environmental conditions, and operator performance.
(2) Monitor over time the accuracy and precision of test performance that may be influenced by changes in test system performance and environmental conditions, and variance in operator performance.
(d) Unless CMS approves a procedure, specified in Appendix C of the State Operations Manual (CMS Pub. 7), that provides equivalent quality testing, the laboratory must—
(1) Perform control procedures as defined in this section unless otherwise specified in the additional specialty and subspecialty requirements at §§ 493.1261 through 493.1278.
(2) For each test system, perform control procedures using the number and frequency specified by the manufacturer or established by the laboratory when they meet or exceed the requirements in paragraph (d)(3) of this section.
(3) At least once each day patient specimens are assayed or examined perform the following for—
(i) Each quantitative procedure, include two control materials of different concentrations;
(ii) Each qualitative procedure, include a negative and positive control material;
(iii) Test procedures producing graded or titered results, include a negative control material and a control material with graded or titered reactivity, respectively;
(iv) Each test system that has an extraction phase, include two control materials, including one that is capable of detecting errors in the extraction process; and
(v) Each molecular amplification procedure, include two control materials and, if reaction inhibition is a significant source of false negative results, a control material capable of detecting the inhibition.
(4) For thin layer chromatography—
(i) Spot each plate or card, as applicable, with a calibrator containing all known substances or drug groups, as appropriate, which are identified by thin layer chromatography and reported by the laboratory; and
(ii) Include at least one control material on each plate or card, as applicable, which must be processed through each step of patient testing, including extraction processes.
(5) For each electrophoretic procedure include, concurrent with patient specimens, at least one control material containing the substances being identified or measured.
(6) Perform control material testing as specified in this paragraph before resuming patient testing when a complete change of reagents is introduced; major preventive maintenance is performed; or any critical part that may influence test performance is replaced.
(7) Over time, rotate control material testing among all operators who perform the test.
(8) Test control materials in the same manner as patient specimens.
(9) When using calibration material as a control material, use calibration material from a different lot number than that used to establish a cut-off value or to calibrate the test system.
(10) Establish or verify the criteria for acceptability of all control materials.
(i) When control materials providing quantitative results are used, statistical parameters (for example, mean and standard deviation) for each batch and lot number of control materials must be defined and available.
(ii) The laboratory may use the stated value of a commercially assayed control material provided the stated value is for the methodology and instrumentation employed by the laboratory and is verified by the laboratory.
(iii) Statistical parameters for unassayed control materials must be established over time by the laboratory through concurrent testing of control materials having previously determined statistical parameters.
(e) For reagent, media, and supply checks, the laboratory must do the following:
(1) Check each batch (prepared in-house), lot number (commercially prepared) and shipment of reagents, disks, stains, antisera, (except those specifically referenced in § 493.1261(a)(3)) and identification systems (systems using two or more substrates or two or more reagents, or a combination) when prepared or opened for positive and negative reactivity, as well as graded reactivity, if applicable.
(2) Each day of use (unless otherwise specified in this subpart), test staining materials for intended reactivity to ensure predictable staining characteristics. Control materials for both positive and negative reactivity must be included, as appropriate.
(3) Check fluorescent and immunohistochemical stains for positive and negative reactivity each time of use.
(4) Before, or concurrent with the initial use—
(i) Check each batch of media for sterility if sterility is required for testing;
(ii) Check each batch of media for its ability to support growth and, as appropriate, select or inhibit specific organisms or produce a biochemical response; and
(iii) Document the physical characteristics of the media when compromised and report any deterioration in the media to the manufacturer.
(5) Follow the manufacturer's specifications for using reagents, media, and supplies and be responsible for results.
(f) Results of control materials must meet the laboratory's and, as applicable, the manufacturer's test system criteria for acceptability before reporting patient test results.
(g) The laboratory must document all control procedures performed.
(h) If control materials are not available, the laboratory must have an alternative mechanism to detect immediate errors and monitor test system performance over time. The performance of alternative control procedures must be documented.
[68 FR 3703, Jan. 24, 2003; 68 FR 50724, Aug. 22, 2003]
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1255 Standard: Calibration and calibration verification procedures.
Calibration and calibration verification procedures are required to substantiate the continued accuracy of the test system throughout the laboratory's reportable range of test results for the test system. Unless otherwise specified in this subpart, for each applicable test system the laboratory must do the following:
(a) Perform and document calibration procedures—
(1) Following the manufacturer's test system instructions, using calibration materials provided or specified, and with at least the frequency recommended by the manufacturer;
(2) Using the criteria verified or established by the laboratory as specified in § 493.1253(b)(3)—
(i) Using calibration materials appropriate for the test system and, if possible, traceable to a reference method or reference material of known value; and
(ii) Including the number, type, and concentration of calibration materials, as well as acceptable limits for and the frequency of calibration; and
(3) Whenever calibration verification fails to meet the laboratory's acceptable limits for calibration verification.
(b) Perform and document calibration verification procedures—
(1) Following the manufacturer's calibration verification instructions;
(2) Using the criteria verified or established by the laboratory under § 493.1253(b)(3)—
(i) Including the number, type, and concentration of the materials, as well as acceptable limits for calibration verification; and
(ii) Including at least a minimal (or zero) value, a mid-point value, and a maximum value near the upper limit of the range to verify the laboratory's reportable range of test results for the test system; and
(3) At least once every 6 months and whenever any of the following occur:
(i) A complete change of reagents for a procedure is introduced, unless the laboratory can demonstrate that changing reagent lot numbers does not affect the range used to report patient test results, and control values are not adversely affected by reagent lot number changes.
(ii) There is major preventive maintenance or replacement of critical parts that may influence test performance.
(iii) Control materials reflect an unusual trend or shift, or are outside of the laboratory's acceptable limits, and other means of assessing and correcting unacceptable control values fail to identify and correct the problem.
(iv) The laboratory's established schedule for verifying the reportable range for patient test results requires more frequent calibration verification.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1281 Standard: Comparison of test results.
(a) If a laboratory performs the same test using different methodologies or instruments, or performs the same test at multiple testing sites, the laboratory must have a system that twice a year evaluates and defines the relationship between test results using the different methodologies, instruments, or testing sites.
(b) The laboratory must have a system to identify and assess patient test results that appear inconsistent with the following relevant criteria, when available:
(1) Patient age.
(2) Sex.
(3) Diagnosis or pertinent clinical data.
(4) Distribution of patient test results.
(5) Relationship with other test parameters.
(c) The laboratory must document all test result comparison activities.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1236 Standard: Evaluation of proficiency testing performance.
(a) The laboratory must review and evaluate the results obtained on proficiency testing performed as specified in subpart H of this part.
(b) The laboratory must verify the accuracy of the following:
(1) Any analyte or subspecialty without analytes listed in subpart I of this part that is not evaluated or scored by a CMS-approved proficiency testing program.
(2) Any analyte, specialty or subspecialty assigned a proficiency testing score that does not reflect laboratory test performance (that is, when the proficiency testing program does not obtain the agreement required for scoring as specified in subpart I of this part, or the laboratory receives a zero score for nonparticipation, or late return of results).
(c) At least twice annually, the laboratory must verify the accuracy of the following:
(1) Any test or procedure it performs that is not included in subpart I of this part.
(2) Any test or procedure listed in subpart I of this part for which compatible proficiency testing samples are not offered by a CMS-approved proficiency testing program.
(d) All proficiency testing evaluation and verification activities must be documented.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
§ 493.1235 Standard: Personnel competency assessment policies.
As specified in the personnel requirements in subpart M, the laboratory must establish and follow written policies and procedures to assess employee and, if applicable, consultant competency.
Official text from the Electronic Code of Federal Regulations (eCFR), current as of November 24, 2026.
Frequently asked questions
Can software replace paper logs?
Yes, if the records are complete, secure, retrievable and show who did and reviewed each task.